Why we need use cascade system for high and low temperature test chambers?
Time:2019-9-19 18:59:14  Browse:times

Why we need use cascade system for high and low temperature test chambers?


Many customers ask why our company's high and low temperature test chambers use double compressors, and some use single compressors. We all know that there are two ways to cool the high and low temperature test chambers: the first one is the single-stage cooling method (that is, the single compressor), and the second is the cascade cooling method (that is, the double compressor). When using a single compressor, when will I use a double compressor? Let's explain it :


1. The lowest temperature that can be achieved with a single compressor is -43 ° C, which is still achievable under no-load conditions. So after adding the load, it is very difficult to achieve this value. Therefore, -40 ° C is the transfer node of the high and low temperature test chamber compressor. A single-stage cooling method is generally adopted above -40 ° C, and a cascade system is generally used below -40 ° C. Another point is that it depends on the volume of the high and low temperature test chamber. Generally, a single-stage cooling method can be used for less than 500L, and a cascade refrigeration system is used for those larger than 500L.


2. If there is a requirement for the temperature rise and fall rate, a cascade refrigeration system is generally used. We all know that the speed of the rapid temperature test chamber can reach 3 ° C, 5 ° C, 10 ° C per minute, etc.


3. The use of a cascade refrigeration system also stabilizes machine performance, saves energy and saves heat, and effectively eliminates heat loss. Therefore, for the experimental load is too large or the heat is too large, it is necessary to use the cascade refrigeration system.


4. Why not all high and low temperature test chambers use cascade refrigeration systems? The cost of double compressor is definitely higher than the cost of a single compressor. Secondly, the equipment manufacturing is certainly more complicated than the single-stage refrigeration system, and the welding process requirements are higher.

So do you knows when to use a cascade refrigeration system and when to use a single-stage refrigeration system now. If you have other questions about the high and low temperature test chamber, you can contact us anytime. We sincerely serve every customer!